Paperless Recorder FX1008-4-2-L
Paperless Recorder FX1000: FX1002、FX1004、FX1006、FX1008、FX1010、FX1012
At YOKOGAWA, we are committed to the "quality first, customer first" principle in all areas of our business, including product design, research and development, and sales and services. The new FX1000 paperless recorder exceeds customer expectations for quality, high performance and capability- at a price that meets the needs of a cost-sensitive market.
Contact us:
Detail:
- Intuitive display, easy operation
- 5.7-inch, high-precision, wide-viewing-angle color TFT LCD
- Many types of displays such as trend, digital, bar graph, overview, alarm, and historical trend
- Remote viewing of the FX1000 screen through the Internet
- Multi-functional panel keys
- Start/stop recording by batch, and create data files
- Comprehensive selection of measurement types, and exceptional performance
- Input types: DCV, TC, RTD, DI
- Scan interval: 1 s, 125 ms (fast sampling)
- Channels: 2, 4, 6, 8, 10, 12
- Measurement accuracy: ±0.05% of reading (DCV), ±0.15% of reading (TC, RTD)
- Reliable data storage
- Large (400 MB) internal memory
- CF cards up to 2 GB (option)
- USB interface (option)
- Binary data storage
- Network enables data redundancy
- Support for a variety of applications
- Power measurement recording (option)
- Vacuum pressure recording (Log scale, option)
- Flow rate summation (option)
- F value calculation (option)
- Reliable design and construction
- Space-saving design
- Waterproof and dustproof (IP65 compliant)
Monitor display
You can use the keys to switch to any of the operation screens. The operating states of memory sampling, alarms, key lock, computation, and other conditions are graphically displayed. Supports Chinese, English, German, French, and Japanese.
Trend display |
Digital display |
Overview display |
Displays measured data as waveforms. Displays each channel's scale value, industrial units, user messages, and other information. |
In addition to displaying measurements digitally, it displays channel/tag, industrial units, and alarm statuses. |
Displays measurements and alarm statuses on all channels. |
Trend display (log scale) |
Stacked bar graph |
|
You can display log scales for degree of vacuum (pressure input), and record the data (option). |
You can display and record power consumption of each piece of electronic equipment; useful for energy saving and equipment maintenance as part of environmental protection programs (option). |
|
Bar graph, historical trend, and information displays (alarm summaries, message summaries, and reports) are also included.
Intuitive operator controls
The DISP/ENTER and arrow keys provide display mode and setting menu navigation. Clearly labeled menu, function, and record start/stop keys handle all setting and control operations.
IE browser monitor screen
Multi-Channel Measurement and Recording
Universal input signal measurement
GENERAL SPECIFICATIONS
Construction
Mounting: |
Flush panel mounting (on a vertical plane)
Mounting may be inclined downward up to 30 degrees from a horizontal plane. |
Allowable panel thickness: |
2 to 26 mm |
Front panel: |
Water- and dustproof: Complies with IEC529-IP65 (excluding side-by-side mounting) |
Input
Number of inputs: |
FX1002: 2 channels, FX1004: 4 channels, FX1006: 6 channels,
FX1008: 8 channels, FX1010: 10 channels, FX1012: 12 channels |
Measurement intervals: |
FX1002, DX1004: 125 ms, 250 ms
FX1006, FX1008, FX1010, FX1012: 1 s, 2 s, 5 s |
Inputs: |
DCV (20, 60, 200 mV, 1, 2, 6, 20, 50 V, 1-5 V)
TC (R, S, B, K, E, J, T, N, W, L, U, WRe)
RTD (Pt100, JPt100)
DI (Contact input, TTL level)
DCA (with external shunt resistor attached) |
Measurement/display accuracy: |
Standard operating conditions: Temperature: 23 ± 2°C
Humidity: 55% ± 10%RH
Power supply voltage: 90 to 132 or 180 to 250 VAC
Power supply frequency: 50/60 Hz ± 1%
Warm-up time: At least 30 minutes. Other ambient conditions such as vibration should not adversely affect the operation. |
Input |
Range |
Measurement Accuracy |
Digital Display Max. Resolution |
DCV |
1-5 V |
±(0.05% of rdg+3 digits) |
1 mV |
Thermocouple* |
K |
±(0.15% of rdg + 0.7°C)
–200 to –100°C: ±(0.15% of rdg + 1°C) |
0.1°C |
RTD |
Pt100 |
±(0.15% of rdg+0.3°C) 0.1°C |
0.1°C |
* Does not include the accuracy of reference junction compensation
Display
Display: |
5.7-inch TFT color LCD (240 × 320 dots)
* A section of the LCD monitor may contain pixels that are always on or off. The brightness of the LCD may also not be uniform due to the characteristics of the LCD. This is not a malfunction. |
Display groups: |
Number of groups: 10
Number of channels that can be assigned to each group: Up to 6 |
Display color: |
Channel: Select from 24 colors
Background: White or black (selectable) |
Trend display: |
Layout: Vertical, horizontal, or wide |
Bar graph display: |
Direction: Vertical or horizontal (selectable) |
Digital display: |
Update rate: 1 sec |
Overview display: |
Measuring values and alarm status of all channels |
Information display: |
Alarm summary, message summary, memory summary, report, stacked bar graph, status, Modbus status |
Modbus log display: |
Displays the login log, error log, communication log (/C2, /C3, and /C7), FTP log (/C7), Web log (/C7), e-mail log (/C7), SNTP log (/C7), and DHCP log (/C7) |
Tag display: |
Number of displayable characters: Up to 16
Displayable characters: English, Japanese, and Chinese |
Messages: |
Number of displayable characters: Up to 32 alphanumeric
Displayable characters: English, Japanese, and Chinese |
Historical display function: |
Plays back data from internal memory or external memory media. |
Back light saver function: |
LCD back light dims or turns OFF (user selectable) if no keys have been pressed for a specified period (1, 2, 5, 10, 30, or 60 min). |
Data Saving Function
External storage medium: |
Medium: CompactFlash memory card (CF card) (on FXs with a CF card slot) |
Internal memory: |
Medium: Flash memory
Format: FAT32 or FAT16
Capacity: 400 MB
Maximum number of files that can be saved: 400 (total number of display data and event data files)
Operation: FIFO (First In, First Out) |
Alarm Function
Number of alarm levels: |
Up to four for each channel |
Alarm types: |
High limit, low limit, differential high limit, difference low limit, high rate-of-change limit, low rate-of-change limit, alarm delay high limit, and alarm delay low limit |
Event Action Function
General: |
A particular action can be executed by particular event. |
Number of event actions: |
40 actions can be set |
Security Functions
General: |
Login function or key lock function can be set for each key operation or communication operation. |
Key lock function: |
On/off and password can be set for each operation key and FUNC operation. |
Login function: |
User name and password to login can be set.
System administrators: 5 (with access to all operations)
Users: 30 (with access to operations based on their user access rights) |
Clock
Clock: |
With calendar function (Western calendar) |
Accuracy: |
±50 ppm (0 to 50°C); does not include the delay (1 second or less) that occurs when the power is turned on. |
Batch Function
General: |
Data display and data management with batch name, text field function and batch comment function are available. |
Power Supply
Rated power supply: |
100 to 240 VAC (automatic switching) |
Allowable power supply voltage range: |
90 to 132 or 180 to 264 VAC |
Rated power supply frequency: |
50/60 Hz (automatic switching) |
Power consumption: |
Max. 45 VA (for 240 VAC power supply) |
NORMAL OPERATING CONDITIONS
Supply voltage: |
90 to 132, 180 to 250 VAC |
Rated power supply frequency: |
50 Hz ±2%, 60 Hz ±2% |
Ambient temperature: |
0 to 50°C |
Ambient humidity: |
20 to 80% RH (at 5 to 40°C), 10 to 50% (at 40 to 50°C) |